List of error codes

 

Er0

The pressure in the air circuit is unstable when the measurement is started.
Try again without moving.

 

Er1

The blood pressure monitor does not detect systolic pressure.

Try again without moving.

 

Er2

The blood pressure monitor does not detect diastolic pressure.

Try again without moving.

 

Er3

The pneumatic system is blocked or the cuff is too tight while inflating.
Reposition the cuff correctly and try again.

 

Er4

Air leak in the pneumatic system or the cuff is not tight.
Reposition the cuff correctly and try again.

 

Er5

The pressure in the cuff is higher than 300 mmHg.

Wait 5 minutes and try again.

 

Er6

The pressure in the cuff is higher than 15 mmHg for more than 3 minutes.

Wait 5 minutes and try again.

 

Er7

Memory fault or error in the temperature probe

Wait 5 minutes and try again.

 

Er8

Abnormality in checking the device's settings.
Wait 5 minutes and try again.

 

Er A

Pressure sensor fault.

Wait 5 minutes and try again.

 

Er

Bluetooth initialisation failed.

Restart your blood pressure monitor and restart the iHealth MyVitals app.

Make sure Bluetooth is activated.

Take a new measurement in 5 minutes.

 

Low battery

The battery is less than 20%.

Recharge the battery.

 

Displayed results seem incorrect

The cuff was positioned incorrectly or the cuff was not tight.
Reposition the cuff correctly and try again.

Body posture was not correct during the test.
Check your posture and try again. 

Please observe the precautions for use carefully.

 

Nothing happens when I press a button

Make sure that the battery is sufficiently charged.

 

 


Your problem persists? Another question? Submit a request