The control solution contains a known quantity of glucose which reacts with the test strips and is used to check that your blood glucose meter and test strips work properly together.
Test results with the control solution must be within the result range printed on the test strip vial.
It is recommended to perform a test with the control solution in the following cases:
- When first unpacking your blood glucose meter.
- Periodic inspection (recommended once a week).
- When opening a new vial of test strips.
- When the blood glucose meter and/or test strips are not operating correctly.
- When your blood glucose level read is not what you expected (according to your general condition / feeling) or you think the result is inaccurate.
- After the tests trips have been exposed to extreme conditions (heat, cold, humidity).
- After dropping your blood glucose meter or a shock.
See the article test with control solution for more information.
Note -The control solution must be used before the expiry date shown on the control solution vial.
Once opened, the control solution must be used within 90 days.
Once opened, the control solution must be used within 90 days.
Comments
0 comments
Please sign in to leave a comment.