Er0
The pressure in the air circuit is unstable when the measurement is started. Try again without moving.
Er1
The blood pressure monitor does not detect systolic pressure.
Try again without moving.
Er2
The blood pressure monitor does not detect diastolic pressure.
Try again without moving.
Er3
The pneumatic system is blocked or the cuff is too tight while inflating.
Reposition the cuff correctly and try again.
Er4
Air leak in the pneumatic system or the cuff is not tight.
Reposition the cuff correctly and try again.
Er5
The pressure in the cuff is higher than 300 mmHg.
Wait 5 minutes and try again.
Er6
The pressure in the cuff is higher than 15 mmHg for more than 3 minutes.
Wait 5 minutes and try again.
Er7
Memory fault or error in the temperature probe.
Wait 5 minutes and try again.
Er8
Abnormality in checking the device's settings.
Wait 5 minutes and try again.
Er A
Pressure sensor fault.
Wait 5 minutes and try again.
The results displayed seem incorrect
Make sure that the monitor is placed correctly.
Place the monitor correctly and try again.
Please observe the precautions for use.
See also:
Nothing happens when I press a button
Check that the battery is sufficiently charged.