E 1
Strip was removed during measurement.
Take a new test with a new test strip.
E 2, E 4
Problem with the reader.
Take a new test with a new test strip.
E 3
Problem with the test strip (wet strip, removed too quickly, too much blood applied to the strip).
Take a new test with a new test strip.
E 5
Ambient temperature too low (<10°C).
E 6
Ambient temperature too high (>40°C).
E 7
Communication error.
Press and hold the M (Memory) button for 7 seconds to restore the glucometer's factory settings.
Restoring factory settings:
- Restores the date by default (01/01/2017). Remember to reconnect the glucometer before taking new offline measurements.
- Do not erase the data in memory.
- Do not delete the glucometer from your list of devices (no need to reinstall the glucometer after reset).
E 8
Result too low (<20 mg/dL).
Take a new test with a new test strip.
E 9
Result too high (>600 mg/dL).
Take a new test with a new test strip.